Technical Specs

  • Absolute Measurement Range
    1.0 x 10-4 to 1000 Torr
  • Digital Communication
    RS485
  • Type
    Vacuum Module
  • Display
    No
  • Interface
    DeviceNet
  • Relays
    2
  • Fitting Type
    8 VCR female
  • Installation Orientation
    Horizontal preferred
  • Set Point Range
    1x10-3 to 1000 Torr, 1x10-3 to 1300 mbar, 1x10-1 Pa to 130 kPa
  • Operating Temperature
    0 - 40°C ambient, non-condensing
  • Maximum Bakeout Temperature
    150°C with electronics removed
  • Storage Temperature
    -40 - 70°C
  • Digital Controls
    Vacuum and atmosphere calibration, set points (value, direction and hysteresis)
  • Analog Output
    0.375-5.659 VDC (0-1000 Torr Non-linear)
  • Relay Resolution
    2 significant digits
  • Power Requirements
    DeviceNet: 11 - 26 VDC, 0.5 Amps at 11 VDC, 2.5 Watts max, analog: 11.5 - 26.5 VDC, 0.1 Amps at 11.5 VDC, 1.6 Watts maximum, 1.8 maximum (with display), RS485: 11.5 - 26.5 VDC, 0.12 Amps at 11.5 VDC, 2 Watts max
  • Internal Volume
    40 cm3 (2.5 in3)
  • Exposed Materials
    304 stainless steel, borosilicate glass, Kovar, alumina, NiFe alloy, polyimide
  • Electronic Casing and Flange
    Aluminum extrusion
  • Compliance
    CE
  • Filament
    Platinum
  • Set Point Contact Rating
    1 Amps at 30 VDC resistive, AC non-inductive
  • Set Point Configuration
    Single-pole, double-throw (SPDT)
  • Set Point Adjustments
    Value, direction, and hysteresis through software
  • Connectors
    9 or 15-pin subminiature-D male, high density: 5-pin Micro DeviceNet
  • Display Resolution
    Least significant digit on each range
  • Units
    Pa
  • Digital Output Resolution
    Least significant digit on each range
  • Digital Baud Rates
    300-115.2K, 19200 Baud (default value)
  • Digital Data Format
    ASCII, 8 data bits, one stop bit, no parity, no handshake (default values)
  • Device Type
    Vacuum/pressure gauge device
  • DeviceNet Messaging
    Polled I/O and explicit
  • DeviceNet Adjustable Parameters
    Vacuum and atmosphere calibrations, set points
  • DeviceNet Address
    0 - 63, selected by using the Low and High address switches
  • DeviceNet Connector
    5-pin Micro
  • DeviceNet Baud Rates
    125K, 250K, or 500K, switch selectable
  • Step Size at Minimum Pressure
    1 x 10-4 Torr, 1 x 10-4 mbar, 1 x 10-2 Pa

Technical Specs

  • Absolute Measurement Range
    1.0 x 10-4 to 1000 Torr
  • Digital Communication
    RS485
  • Type
    Vacuum Module
  • Display
    No
  • Interface
    DeviceNet
  • Relays
    2
  • Fitting Type
    8 VCR female
  • Installation Orientation
    Horizontal preferred
  • Set Point Range
    1x10-3 to 1000 Torr, 1x10-3 to 1300 mbar, 1x10-1 Pa to 130 kPa
  • Operating Temperature
    0 - 40°C ambient, non-condensing
  • Maximum Bakeout Temperature
    150°C with electronics removed
  • Storage Temperature
    -40 - 70°C
  • Digital Controls
    Vacuum and atmosphere calibration, set points (value, direction and hysteresis)
  • Analog Output
    0.375-5.659 VDC (0-1000 Torr Non-linear)
  • Relay Resolution
    2 significant digits
  • Power Requirements
    DeviceNet: 11 - 26 VDC, 0.5 Amps at 11 VDC, 2.5 Watts max, analog: 11.5 - 26.5 VDC, 0.1 Amps at 11.5 VDC, 1.6 Watts maximum, 1.8 maximum (with display), RS485: 11.5 - 26.5 VDC, 0.12 Amps at 11.5 VDC, 2 Watts max
  • Internal Volume
    40 cm3 (2.5 in3)
  • Exposed Materials
    304 stainless steel, borosilicate glass, Kovar, alumina, NiFe alloy, polyimide
  • Electronic Casing and Flange
    Aluminum extrusion
  • Compliance
    CE
  • Filament
    Platinum
  • Set Point Contact Rating
    1 Amps at 30 VDC resistive, AC non-inductive
  • Set Point Configuration
    Single-pole, double-throw (SPDT)
  • Set Point Adjustments
    Value, direction, and hysteresis through software
  • Connectors
    9 or 15-pin subminiature-D male, high density: 5-pin Micro DeviceNet
  • Display Resolution
    Least significant digit on each range
  • Units
    Pa
  • Digital Output Resolution
    Least significant digit on each range
  • Digital Baud Rates
    300-115.2K, 19200 Baud (default value)
  • Digital Data Format
    ASCII, 8 data bits, one stop bit, no parity, no handshake (default values)
  • Device Type
    Vacuum/pressure gauge device
  • DeviceNet Messaging
    Polled I/O and explicit
  • DeviceNet Adjustable Parameters
    Vacuum and atmosphere calibrations, set points
  • DeviceNet Address
    0 - 63, selected by using the Low and High address switches
  • DeviceNet Connector
    5-pin Micro
  • DeviceNet Baud Rates
    125K, 250K, or 500K, switch selectable
  • Step Size at Minimum Pressure
    1 x 10-4 Torr, 1 x 10-4 mbar, 1 x 10-2 Pa

Need help?

Contact an Applications Specialist by sending us an email