MultiGas 2031 Purity FTIR Gas Analyzer
Model: MULTIGAS-2031
MultiGas 2031 Purity FTIR Gas Analyzer
Model: MULTIGAS-2031

Overview
Overview
MultiGas 2031 Purity FTIR Gas Analyzer, for Real-Time Trace Impurity Detection
The MultiGas™ FTIR Gas Analyzer, Model 2031, is an FTIR based analyzer that ensures high gas purity by detecting ppb level contaminants in bulk process gases such as NH3, NF3, and H2. The analyzer uses an FTIR Spectrometer, high-optical throughput sampling cell, and application specific analysis software to ensure gas purity. These integrated components ensure high precision and accuracy for bulk gas contaminant monitoring. Multiple detector and gas cell options are available which allow the analyzer to be tailored to the process.
Applications
Continuous monitoring of trace contaminants in high purity bulk gases in production environments, including:
- H2O, CO, CO2, N2O and CH4 in NH3
- H2O, CO, CO2, N2O and HF in NF3
- H2O, CO, CO2, HF, HCI, NH3, CH2CI2 and CHCI3 in H2
- H2O, CO, CO2, CH4, and NH3 in N2O
Technical Specs
- TypeFTIR Analyzer
- Measurement TechniqueFTIR Spectrometry
- Measurable Gases & VaporsMost molecules except for N2, H2, and O2
- Measurable RangeFull scale concentration setting between 10ppb and 100% Full Scale Spectral Resolution
- Spectral Resolution0.5 - 128cm-1
- Scan Time1-300 sec
- DetectorLN2-cooled MCT, TE-cooled MCT
- Purge Pressure20 psig (1.5 bar) max.
- Spectrometer Purge Flow4 L/min of purified ultra pure N2
- Optics Purge Flow4 L/min of purified ultra pure N2
- Pressure TransducerMKS Baratron® capacitance manometer
- Purge Connection0.25 in. VCR® type
- Computer RequirementsDesktop or Laptop with Windows 10 OS
- CommunicationEthernet
- Enclosure19 in. rack mount chassis
- Dimensions17.5 x 12.5 x 25.5 in. (width, height, depth)
- Power Requirements120 or 240 VAC, 50/60 Hz, 6 amps
- Weight125 lbs. (57 kg)
- Sample TemperatureAmbient to 60°C (calibration temperature dependant)
- Sample Flow0.2 - 10 L/min
- Sample Pressure0.01-8 atm (calibration pressure dependant)
- Laser SafetyClass 1 laser product contains a Class 3R laser with continuous wave output at 633 nm
Technical Specs
- TypeFTIR Analyzer
- Measurement TechniqueFTIR Spectrometry
- Measurable Gases & VaporsMost molecules except for N2, H2, and O2
- Measurable RangeFull scale concentration setting between 10ppb and 100% Full Scale Spectral Resolution
- Spectral Resolution0.5 - 128cm-1
- Scan Time1-300 sec
- DetectorLN2-cooled MCT, TE-cooled MCT
- Purge Pressure20 psig (1.5 bar) max.
- Spectrometer Purge Flow4 L/min of purified ultra pure N2
- Optics Purge Flow4 L/min of purified ultra pure N2
- Pressure TransducerMKS Baratron® capacitance manometer
- Purge Connection0.25 in. VCR® type
- Computer RequirementsDesktop or Laptop with Windows 10 OS
- CommunicationEthernet
- Enclosure19 in. rack mount chassis
- Dimensions17.5 x 12.5 x 25.5 in. (width, height, depth)
- Power Requirements120 or 240 VAC, 50/60 Hz, 6 amps
- Weight125 lbs. (57 kg)
- Sample TemperatureAmbient to 60°C (calibration temperature dependant)
- Sample Flow0.2 - 10 L/min
- Sample Pressure0.01-8 atm (calibration pressure dependant)
- Laser SafetyClass 1 laser product contains a Class 3R laser with continuous wave output at 633 nm
Resources & Downloads
Literature
MultiGas™ 2031 FTIR Gas Analyzer(2 MB, PDF)