Sorry, this product is no longer offered for sale. The recommended replacement product is da02b-ethercat-capacitance-manometers.
The EtherCAT® Baratron® capacitance manometer uses industry-standard EtherCAT digital communications to connect to complex process tools, based on the proven 627C and 628C Ethernet-equipped Baratron products.
Specifications
- TypeDigital Pressure Transducer
- Sensor TypeCapacitance Manometer
- Measurement TypeAbsolute
- Sensor TemperatureControlled to 45°C, 80°C or 100°C
- Full Scale Pressure Ranges0.1, 0.25, 1, 2, 10, 20, 50, 100, 200, 500, and 1000 Torr (and metric equivalents)
- Accuracy45°C Models: 0.10% of Reading for ranges ≥1 Torr, 0.15% Reading for ranges <1 Torr
80° and 100°C Models: 0.25% of Reading for ranges ≥1 Torr, 0.50% of Reading for ranges <1 Torr - Resolution0.001% Full Scale
- Communication ProtocolEtherCAT®
- Analog Output0 - 10 VDC into >10Ω load
- Analog Connector15-pin D-subminiature male
- Digital ConnectorTwo (2) RJ45 female receptacles for incoming and network signals
- Zero Temperature Coefficient45°C Models:0.002% Full Scale/°C for ranges ≥1 Torr; 0.005% Full Scale/°C for ranges <1 Torr
80° and 100°C Models: 0.002% Full Scale/°C for ranges ≥1 Torr; 0.010% Full Scale/°C for ranges <1 Torr - Span Temperature Coefficient0.02% Reading/°C
- Ambient Operating Temperature45°C Models: 15° to 40°C
80° and 100°C Models: 15° to 50°C - Volume6.3 cc
- Warm-up Time2 hours for ranges ≥1 Torr
4 hours for ranges <1 Torr - Overpressure Limit45 psia (310 kPa)
- Materials Exposed to Process GasesInconel® and Incoloy® nickel alloys
- Power Requirements+24 VDC ±10% (Unheated 300mA, 45°C Models 600 mA, 80° and 100°C Models 800 mA)
±15VDC ±5% (Unheated 300mA, 45°C Models 600 mA, 80° and 100°C Models 800 mA) - Compliance2014/30/EU, 2011/65/EU
Accessories
Power Supplies and Displays
| Compare | Description | Drawings, CAD & Specs | Avail. | Price | ||
|---|---|---|---|---|---|---|
![]() | PR4000BF2V2Digital Power Supply, Display, 2 Channel, RS232, ±15VDC, 1.5 Amp | In Stock | ||||
![]() | PR4000BF2V3Digital Power Supply, Display, 2 Channel, RS232, ±24VDC, 1.0 Amp | |||||
![]() | PR4000BF5V2Digital Power Supply, Display, 2 Channel, RS485, ±15VDC, 1.5 Amp | |||||
![]() | PR4000BS2V3Digital Power Supply, Display, 1 Channel, RS485, ±24VDC, 1.0 Amp | |||||
![]() | PR4000BF5V3Digital Power Supply, Display, 2 Channel, RS485, ±24VDC, 1.0 Amp | |||||
![]() | PR4000BS2V2Digital Power Supply, Display, 1 Channel, RS232, ±15VDC, 1.5 Amp | |||||
![]() | PDR2000AODigital Power Supply, Display, 2 Channels, RS232, ±15 VDC, 0.75 Amp | In Stock |
Resources
Device Description Files
DA02A Baratron® Capacitance Manometer ESI File(171.1 kB, XML)
Drawings & CADs
DA02 Baratron® Capacitance Manometer Dimensional Drawings(131.9 kB, PDF)
Literature
DA02A EtherCAT® Baratron® Capacitance Manometer (archive)(259.9 kB, PDF)
Manuals
DA02 Baratron® Capacitance Manometer with EtherCAT® Pin-outs(23.9 kB, PDF) DA02A EtherCAT Manometer Manual(612.9 kB, PDF)
Application Notes
Minimizing Fluorine-Induced Drift in Capacitance Manometers(307.1 kB, PDF)