Compare Description Drawings, CAD & Specs Availability Price
600-series heated absolute baratron capacitance manometer
Heated Capacitance Manometer, 0.1 Torr, 45°C, 8 VCR-F, DB9 w/Thread Lock, 0.25% of Rdg Accuracy, Status LED, External Zero
600-series heated absolute baratron capacitance manometer
Heated Capacitance Manometer, 0.1 Torr, 45°C, 8 VCR-F, DB9 w/Thread Lock, 0.50%/Rdg Accuracy, Status LED
600-series heated absolute baratron capacitance manometer
Heated Capacitance Manometer, 10 Torr, 45°C, 8 VCR-F, DB9 w/Thread Lock, 0.12% Rdg Accuracy, Status LED
$2,399
600-series heated absolute baratron capacitance manometer
Heated Capacitance Manometer, 0.1 Torr, 100°C, 8 VCR-F, DB9 w/Thread Lock, 0.25%/Rdg Accuracy
600-series heated absolute baratron capacitance manometer
Heated Capacitance Manometer, 10 Torr, 45°C, NW16 ISO-KF, DB9 w/Thread Lock, 0.12%/Rdg Accuracy, Status LED, External Zero
$2,454
600-series heated absolute baratron capacitance manometer
Heated Capacitance Manometer, 10 Torr, 100°C, NW16 ISO-KF, DB9 w/Thread Lock, 0.25%/Rdg Accuracy, External Zero
600-series heated absolute baratron capacitance manometer
Heated Capacitance Manometer, 1,000 Torr, 45°C, 8 VCR-F, DB9 w/Thread Lock, 0.12% Rdg Accuracy
$2,288
600-series heated absolute baratron capacitance manometer
Heated Capacitance Manometer, 1,000 Torr, 100°C, 8 VCR-F, DB9 w/Thread Lock, 0.12% Rdg Accuracy
600-series heated absolute baratron capacitance manometer
Heated Capacitance Manometer, 1,000 Torr, 45°C, NW16 ISO-KF, DB9 w/Thread Lock, 0.12% Rdg Accuracy, Status LED
$2,371
600-series heated absolute baratron capacitance manometer
Heated Capacitance Manometer, 1,000 Torr, 100°C, NW16 ISO-KF, DB9 w/Thread Lock, 0.12% Rdg Accuracy

Configuration Options

Description
600-series heated absolute baratron capacitance manometer 629H Absolute Capacitance Manometer, 0.02-20,000 Torr, 9-Pin D-Sub

Specifications

  • Type
    Vacuum/pressure Transducer
  • Measurement Type
    Absolute
  • Sensor Temperature
    Controlled to 45°C or 100°C
  • Full Scale Pressure Ranges
    0.02 (45°C only), 0.05, 0.1, 0.25, 1, 2, 10, 20, 100, 1000, 2000, 5000, 10000, 15000, 20000, 25000 Torr
  • Accuracy
    45°C version: 0.12% of reading for 1-25,000 Torr
    0.15% of reading for 0.25, 0.1, and 0.05 Torr
    0.25% for 0.02 Torr
  • Analog Output
    0-10 VDC into >10K Ω load
  • Analog Connector
    9-pin D-sub
  • Span Temperature Coefficient
    0.02% of Reading/°C
  • Operating Temperature Range
    45°C version: 15-40°C
    100°C version: 15-50°C
  • Volume
    6.3 cc
  • Warm-up Time
    2 hours for 1-25,000 Torr ranges, 4 hours for ranges < 1 Torr
  • Overpressure Limit
    45 psia (310 kPa) or 120% Full Scale, whichever is greater
  • Exposed Materials
    Inconel®
  • Power Requirements
    ±15 VDC ±5% @ 0.25 A (45°C) or 0.6 (100°C)
  • Compliance
    CE

Features

Baratron® Capacitance Manometer Technology

Capacitance manometers are electro-mechanical gauges that can measure both pressure and vacuum. The capacitance gauge translates a pressure-modulated movement in a thin diaphragm into an electrical signal proportional to the pressure. The pressure sensor is the thin diaphragm that is exposed to the pressure or vacuum being measured via the inlet tube. An electrode is mounted in the reference cavity behind the diaphragm. Pressure differences between the process and the reference cavity deflect the diaphragm slightly, changing the distance between it and the electrode. Variations in this distance produce variations in the capacitance between the diaphragm and the electrode creating an electrical signal that is proportional to the pressure change. Since differences in the capacitance signal are produced by physical changes within the manometer and not by changes in the gas properties, pressure measurements by the capacitance manometer are independent of the composition of the gas being measured.

Gas independent pressure measurement with a Baratron® capacitance manometer. Schematic shows the internal components and functional zones.

Percent of Reading Accuracy

Accuracy is specified as a percent of Reading, not Full Scale, as seen in some of the lower performance devices. Percent of Reading accuracy provides you with an even more accurate output signal in the lower scale of the pressure range, where it is needed most.

Internally Heated to 45°C or 100°C

These capacitance manometers are temperature controlled to 45°C or 100°C for improved accuracy. Unheated versions are exposed to ambient temperature variations which can degrade the sensor accuracy. These devices have the sensor enclosed in a volume that is maintained at a constant temperature above ambient. This solution improves the manometer’s accuracy and repeatability and lowers instrument drift by reducing or eliminating process contamination within the manometer. Heated manometers are recommended for applications that require maximum accuracy and repeatability, operate above ambient temperature and for those processes that employ hot gases.

The pressure sensor is located inside a 45°C or 100°C constant temperature oven for improved pressure measurement accuracy.

Absolute Pressure Measurement

These Baratron® pressure transducers are referenced to vacuum for absolute pressure measurement. Applications include: vacuum furnaces, freeze-drying of fruits and vegetables, gas lasers, automotive component testing, bottle coatings, and vacuum distillation.

All-Inconel® Wetted Surfaces

These pressure transducers feature all Inconel® construction of the pressure sensor allowing it to operate without damage in virtually any chemical environment, including exposure to halogens, deionized water and steam, and ozone.

0 to 10 VDC Proportional Analog Output

These Baratron® capacitance manometers feature a high-level 0-10 VDC analog output signal that is linear with pressure. This analog output can be interfaced with an MKS pressure controller, an MKS power supply/display instrument, or any instrument that meets these requirements.

9-pin Type "D" Pin-outs

Pin Description
1 Pressure Output
2 No Connection
3 No Connection
4 +15 VDC
5 -15 VDC
6 No Connection
7 External Zero
8 Pressure Common
9 Power Common

Need help?

Contact an Applications Specialist by sending us an email