The 626D Unheated Absolute Baratron® Capacitance Manometer is the latest of MKS analog capacitance-based, high-performance vacuum and pressure transducers. It requires ±15VDC input voltage and provides a high-level 0-10 VDC analog output signal that is linear with pressure.
Compare | Description | Drawings, CAD & Specs | Availability | Price | |||
---|---|---|---|---|---|---|---|
626D01TBE Absolute Pressure Transducer, 1 Torr, 0-10 VDC, 8 VCR-F, 15-pin D-sub | $1,390 |
|
|||||
Absolute Pressure Transducer, 1 Torr, 0-10 VDC, NW16 ISO-KF, 15-pin D-sub
$1,390
|
626D01TDE Absolute Pressure Transducer, 1 Torr, 0-10 VDC, NW16 ISO-KF, 15-pin D-sub | $1,390 |
|
||||
626D02TBE Absolute Pressure Transducer, 2 Torr, 0-10 VDC, 8 VCR-F, 15-pin D-sub | $1,390 |
|
|||||
Absolute Pressure Transducer, 2 Torr, 0-10 VDC, NW16 ISO-KF, 15-pin D-sub
$1,390
|
626D02TDE Absolute Pressure Transducer, 2 Torr, 0-10 VDC, NW16 ISO-KF, 15-pin D-sub | $1,390 |
|
||||
626D11TBE Absolute Pressure Transducer, 10 Torr, 0-10 VDC, 8 VCR-F, 15-pin D-sub | $1,390 |
|
|||||
Absolute Pressure Transducer, 10 Torr, 0-10 VDC, NW16 ISO-KF, 15-pin D-sub
$1,390
|
626D11TDE Absolute Pressure Transducer, 10 Torr, 0-10 VDC, NW16 ISO-KF, 15-pin D-sub | $1,390 |
|
||||
626D12TBE Absolute Pressure Transducer, 100 Torr, 0-10 VDC, 8 VCR-F, 15-pin D-sub | $1,390 |
|
|||||
Absolute Pressure Transducer, 100 Torr, 0-10 VDC, NW16 ISO-KF, 15-pin D-sub
$1,390
|
626D12TDE Absolute Pressure Transducer, 100 Torr, 0-10 VDC, NW16 ISO-KF, 15-pin D-sub | $1,390 |
|
||||
Absolute Pressure Transducer, 1000 Torr, 0-10 VDC, 8 VCR-F, 15-pin D-sub
$1,390
|
626D13TBE Absolute Pressure Transducer, 1000 Torr, 0-10 VDC, 8 VCR-F, 15-pin D-sub | $1,390 |
|
||||
Absolute Pressure Transducer, 1000 Torr, 0-10 VDC, NW16 ISO-KF, 15-pin D-sub
$1,390
|
626D13TDE Absolute Pressure Transducer, 1000 Torr, 0-10 VDC, NW16 ISO-KF, 15-pin D-sub | $1,390 |
|
The following options are available for 626D Baratron® Absolute Capacitance Manometers
Ordering Code Example: 626DXXXYZ
Configuration Option | Option Code |
---|---|
626D Absolute Capacitance Manometer | 626D |
Full Scale Range |
|
0.1 Torr | .1T |
0.25 Torr | RET |
1 Torr | 01T |
2 Torr | 02T |
10 Torr | 11T |
20 Torr | 21T |
100 Torr | 12T |
500 Torr | 52T |
1000 Torr | 13T |
Fittings |
|
0.5 in. (12.7 mm) OD tube | A |
8 female VCR | B |
1.33 in. OD (33.8 mm) Conflat, rotatable | C |
NW16 ISO-KF | D |
8 female VCO | E |
NW25 ISO-KF | Q |
Accuracy |
|
0.25% Reading (1 to 1000 Torr) - standard | E |
0.15% Reading (10 to 1000 Torr) - optional | D |
0.50% of Reading (< 1 Torr) - standard | F |
Capacitance manometers are electro-mechanical gauges that can measure both pressure and vacuum. The capacitance gauge translates a pressure-modulated movement in a thin diaphragm into an electrical signal proportional to the pressure. The pressure sensor is the thin diaphragm that is exposed to the pressure or vacuum being measured via the inlet tube. An electrode is mounted in the reference cavity behind the diaphragm. Pressure differences between the process and the reference cavity deflect the diaphragm slightly, changing the distance between it and the electrode. Variations in this distance produce variations in the capacitance between the diaphragm and the electrode creating an electrical signal that is proportional to the pressure change. Since differences in the capacitance signal are produced by physical changes within the manometer and not by changes in the gas properties, pressure measurements by the capacitance manometer are independent of the composition of the gas being measured.
These Baratron® pressure transducers are referenced to vacuum for absolute pressure measurement. Applications include: vacuum furnaces, freeze-drying of fruits and vegetables, gas lasers, automotive component testing, bottle coatings, and vacuum distillation.
These pressure transducers feature Inconel® and Incoloy® nickel alloy construction of the pressure sensor allowing it to operate without damage in virtually any chemical environment, including halogens, deionized water and steam, and ozone. The sensor is fully welded with no intermediate brazing or joining materials.
These Baratron® capacitance manometers feature a high-level 0-10 VDC analog output signal that is linear with pressure. This analog output can be interfaced with an MKS pressure controller, an MKS power supply/display instrument, or any instrument that meets these requirements. They are powered by a ±15 VDC input voltage.
Pin | Description |
---|---|
2 | Signal Output |
5 | Power Common |
6 | -15 VDC |
7 | +15 VDC |
12 | Signal Common |
15 | Chassis Ground |