Technical Specs
Transfer Standard
AA06A Baratron capacitance manometer absolute pressure sensor (NIST traceable)
Resolution
1 x 10-6
Operating Temperature
15-40°C, temperature-controlled to 45°C
Readout
670C, 5½-digit LCD
Power Requirements
110 VAC, 60 Hz; 220 VAC, 50 Hz; 100 VAC, 50/60 Hz; 3A
Warm-up Time
4 hours. Use of the internal 280 VA UPS will keep the transfer standard warmed up and ready for immediate use for at least 60 minutes.
Enclosure
Metal wheeled cart with extendable handle
Weight
75 lb. (34 Kg)
Full Scale Range
0.1, 1, 10, 100, 1000, 5000, 10000, 15000, 20000, 25000 Torr
Configuration
Removable transfer standard (13 ft cable with 4 VCR female transducer fitting)
Baratron Transducer Accuracy
Standard: 0.05% of Reading (±temp. coefficients) - (for 1-1000 range)
Optional: 0.08% of Reading (±temp. coefficients)
Technical Specs
Transfer Standard
AA06A Baratron capacitance manometer absolute pressure sensor (NIST traceable)
Resolution
1 x 10-6
Operating Temperature
15-40°C, temperature-controlled to 45°C
Readout
670C, 5½-digit LCD
Power Requirements
110 VAC, 60 Hz; 220 VAC, 50 Hz; 100 VAC, 50/60 Hz; 3A
Warm-up Time
4 hours. Use of the internal 280 VA UPS will keep the transfer standard warmed up and ready for immediate use for at least 60 minutes.
Enclosure
Metal wheeled cart with extendable handle
Weight
75 lb. (34 Kg)
Full Scale Range
0.1, 1, 10, 100, 1000, 5000, 10000, 15000, 20000, 25000 Torr
Configuration
Removable transfer standard (13 ft cable with 4 VCR female transducer fitting)
Baratron Transducer Accuracy
Standard: 0.05% of Reading (±temp. coefficients) - (for 1-1000 range)
Optional: 0.08% of Reading (±temp. coefficients)