Configuration Options

Description
DA06A Manometer, 3-Decade Temperature-Controlled EtherCAT®

Specifications

  • Type
    Temperature-Controlled Capacitance Manometer
  • Accuracy
    1000 mT to 10mT: ±0.4% of Reading
    8 mT: ±0.5% of Reading
    5 mT: ±0.7% of Reading
    3 mT: ±0.8% of Reading
    2 mT: ±1.0% of Reading
    1 mT3: ±1.5% of Reading
  • Response Time
    Pressure: 12 ms nominal (10-90% of a step response) influenced by step size and digital pressure filter setting
    EtherCAT: Refresh Rate 1 ms
  • Fitting Type
    0.50 in. OD tube, 8 VCR Female, 8 VCR Male, NW16-KF, NW25-KF, Mini-CF (1.33 in.)
    Contact MKS if alternative fitting is required.
  • Output Signal
    EtherCAT®
  • Electrical Connector
    Power: 15-pin D-subminiature male
    EtherCAT: Two RJ45 female receptacles for incoming and network signals
  • Zero Temperature Coefficient
    0.002% Full Scale/°C
  • Span Temperature Coefficient
    0.02% Reading/°C
  • Ambient Operating Temperature
    15° to 60°C (15-40 for 45°C manometer)
  • Volume
    6.3 cm3 typical.
    Contact MKS for volumes with specific flanges.
  • Warm-up Time
    4 hours
  • Overpressure Limit
    45 psia (310 kPA)
  • Materials Exposed to Process Gases
    Inconel® and Incoloy® nickel alloys (Fittings 300 Series Stainless Steel)
  • Power Requirements
    +24 VDC ±10% @ 700 mA max
    ±15 VDC ±5% @ 700 mA max
  • Indicators
    Green RUN LED and Red ERR LED for EtherCAT and Manometer status
    Green Link/Activity Indicator on bottom right of RJ45 connectors
    Rotary hex switches supporting configured station alias and explicit device identification
  • Compliance
    CE, UKCA

Need help?

Contact an Applications Specialist by sending us an email