Differential Pressure Transducer, 45C, 100 Torr, 0.08%, Unidirectional

Technical Specs

  • Type
    Vacuum Pressure Transducer
  • Sensor Type
    Reference Capacitance Manometer
  • Measurement Type
    Differential
  • Full Scale Range
    100 Torr
  • Sensor Temperature
    45°C
  • Fitting Type
    4 VCR female
  • Accuracy
    0.08% of Reading
  • Resolution
    1x10-6 of Full Scale
  • Calibration
    Unidirectional
  • Electrical Connector
    15-pin D-sub
  • Zero Temperature Coefficient
    <0.002% Reading/°C (20 ppm)
  • Span Temperature Coefficient
    <4 PPM, Full Scale/°C
  • Maximum Line Pressure
    150 psig
  • Reference Side (PR) Media Compatibility
    Any dry gas compatible with Inconel®, 304/306 SS, Fosterite, Palladium
  • Measurement Side (Px) Media Compatibility
    Any gas compatible with Inconel®, 304 stainless steel
  • Px Overpressure
    45 psia
  • Volume
    Measurement (Px): 3.5 cc
    Reference (Pr): 25 cc
  • Time Constant with Signal Conditioner
    Fast Position: >25 ms, Standard Position: 40 ms, Slow Position: 400 ms
  • Line Pressure Effects on Span
    <0.010%/15 psi3