Product Comparison

Choose products to compare anywhere you see 'Add to Compare' or 'Compare' options displayed.

Sorry, this product is no longer offered for sale. The recommended replacement product is AD06A.

Technical Specs

  • Full Scale Range
    1 Torr
  • Sensor Temperature
    45°C
  • Fitting Type
    4 VCR female
  • Accuracy
    ±0.12% of Reading Unidirectional
  • Resolution
    1 x 10-6 of Full Scale
  • Output Signal
    Unidirectional: 0 - 10 VDC into >10 kOhm load, Bidirectional: -10 - 10 VDC into >10 kOhm load
  • Electrical Connector
    25-pin type D, male
  • Zero Temperature Coefficient
    12 ppm/°C
  • Span Temperature Coefficient
    100 ppm/°C
  • Operating Temperature
    15 - 40°C
  • Maximum Line Pressure
    kPa (150 psig)
  • Exposed Materials
    Measurement (Px): Inconel, 304 & 316 stainless steel, Reference (Pr): Inconel, 304 & 316 stainless steel, ceramic, Palladium Palladium
  • Px Overpressure
    125% of Full Scale or 35 psig whichever is larger: Pr>Px: 125% of Full Scale
  • Volume
    Measurement (Px): 3.5 cc, 14 cc (unidirectional only), Reference (Pr):25.0 cc, 8.0 cc (unidirectional only)
  • Response Time
    <40 msec
  • Warm-up Time
    4 hours
  • Power Requirements
    ±15 or +24 to 30 VDC @ 700 mA, on turn-on,450 mA after warm-up
  • Range Multiplier
    x1 or x0.1 ranges selectable via TTL, low or pin grounding
  • Remote Zero Adjust
    Zero adjust initiated via TTL low or pin grounding adjustment range: 2% of Full Scale

Technical Specs

  • Full Scale Range
    1 Torr
  • Sensor Temperature
    45°C
  • Fitting Type
    4 VCR female
  • Accuracy
    ±0.12% of Reading Unidirectional
  • Resolution
    1 x 10-6 of Full Scale
  • Output Signal
    Unidirectional: 0 - 10 VDC into >10 kOhm load, Bidirectional: -10 - 10 VDC into >10 kOhm load
  • Electrical Connector
    25-pin type D, male
  • Zero Temperature Coefficient
    12 ppm/°C
  • Span Temperature Coefficient
    100 ppm/°C
  • Operating Temperature
    15 - 40°C
  • Maximum Line Pressure
    kPa (150 psig)
  • Exposed Materials
    Measurement (Px): Inconel, 304 & 316 stainless steel, Reference (Pr): Inconel, 304 & 316 stainless steel, ceramic, Palladium Palladium
  • Px Overpressure
    125% of Full Scale or 35 psig whichever is larger: Pr>Px: 125% of Full Scale
  • Volume
    Measurement (Px): 3.5 cc, 14 cc (unidirectional only), Reference (Pr):25.0 cc, 8.0 cc (unidirectional only)
  • Response Time
    <40 msec
  • Warm-up Time
    4 hours
  • Power Requirements
    ±15 or +24 to 30 VDC @ 700 mA, on turn-on,450 mA after warm-up
  • Range Multiplier
    x1 or x0.1 ranges selectable via TTL, low or pin grounding
  • Remote Zero Adjust
    Zero adjust initiated via TTL low or pin grounding adjustment range: 2% of Full Scale

Need help?

Contact an Applications Specialist by sending us an email