Sorry, this product is no longer offered for sale. The recommended replacement product is aa06a.

Reference Capacitance Manometer, Absolute, 1000 Torr, 0.12% Accuracy, 45°C, Signal Conditioner

Technical Specs

  • Sensor Temperature
    45°C
  • Full Scale Range
    1000 Torr
  • Fitting Type
    4 VCR female
  • Output Signal
    0 - 10 VDC into >10 kOhm load
  • Electrical Connector
    25-pin type D, male
  • Accuracy
    ±0.12% of Reading
  • Resolution
    1x10-6 of Full Scale
  • Zero Temperature Coefficient
    15 ppm/°C
  • Span Temperature Coefficient
    100 ppm/°C
  • Operating Temperature
    15 - 40°C
  • Exposed Materials
    Inconel, 304 & 316 stainless steel
  • Volume
    3.5 cc, 14 cc
  • Warm-up Time
    4 hours
  • Time Response
    <40 msec
  • Power Requirements
    ±15 or +24 to 30 VDC @ 700 mA on turn-on,450 mA after warm-up
  • Px Overpressure
    125% of Full Scale or 35 psia whichever is larger
  • Range Multiplier
    x1 or x0.1 ranges selectable via TTL, low or pin grounding
  • Remote Zero Adjust
    Zero adjust initiated via TTL low or pin grounding adjustment range: 2% of Full Scale

Need help?

Contact an Applications Specialist by sending us an email