Sorry, this product is no longer offered for sale. It was discontinued on Nov 17, 2023
1152C Mass-Flo® Vapor Source Mass Flow Controller with Viscous Laminar Flow
Model: 1152C
1152C Mass-Flo® Vapor Source Mass Flow Controller with Viscous Laminar Flow
Model: 1152C

Overview
Overview
The 1152C Vapor Source Mass Flow Controller is a pressure-based measurement and control system designed to meter and control vapor from low vapor pressure liquid and solid sources directly, without the need of a carrier gas. The 1152C consists of a fixed flow element and two capacitance manometers for flow measurement, with a proportioning solenoid control valve for flow control (U.S. Patent No. 4,679,585). The 1152C contained within a compact temperature-controlled assembly with a temperature status LED and relay to indicate when temperature is in range. The 1152C Heated Pressure-Based Mass Flow Controller using viscous flow through a laminar flow tube is capable of delivering vaporized liquid source materials such as: TEOS, DADBS, HMDS, TMCTS, TEAL, TEB, TEG, TEI, TMAL, TMB, TMG, TMI, TaCl5, DMEAA, It[OCH(CH3)2]4, TiCl4, TIBAL, and TMP.
Technical Specs
- TypeVapor Mass Flow Controller
- Flow Methodviscous flow through a laminar flow tube
- Full Scale Flow Range1 - 1000 sccm (N2 equivalent typical - dependent upon source material)
- Control Range5 - 100% of Full Scale
- Typical Accuracy±5.0% of Full Scale
- Repeatability±0.2% of Full Scale
- Resolution±0.1% of Full Scale
- Operating Temperature30 - 100°C, adjustable
- Optional Operating Temperature Range90 - 150°C
- Settling Time1 second to within 2% of set point
- Controller Input Power Required±15 VDC (±2%) @ 0.28 Amps
- Heater Input Power Required±15 VDC @ 1.5 Amps (3 Amps with 90 - 150°C option)
- Set Point Signal0 - 5 VDC from <20 kOhm
- Flow Output Signal0 - 5 VDC into >10 kOhm
- Temperature Output10 mV/°C (0°C=0 VDC)
- Relay Contact Ratings2 Amps @ 28 VDC, 1 Amp @ 120 VAC resistive
- Signal Connector15-pin Type "D", RFI/EMI shielded
- Heater Connector9-pin Type "D", RFI/EMI shielded
- Maximum Line Pressure35 psia
- External Leak Integrity<1 x 10-9 scc/sec He
- Leak Integrity Through Closed Valve<1% of Full Scale or 1 sccm, whichever is greater
- Wetted MaterialsInconel, 316 Stainless Steel, Nickel
- Mounting PositionDo not mount upside-down
- Fitting TypeSwagelok® 8 VCR® male
- Weight7.9 lbs (3.59 kg)
Technical Specs
- TypeVapor Mass Flow Controller
- Flow Methodviscous flow through a laminar flow tube
- Full Scale Flow Range1 - 1000 sccm (N2 equivalent typical - dependent upon source material)
- Control Range5 - 100% of Full Scale
- Typical Accuracy±5.0% of Full Scale
- Repeatability±0.2% of Full Scale
- Resolution±0.1% of Full Scale
- Operating Temperature30 - 100°C, adjustable
- Optional Operating Temperature Range90 - 150°C
- Settling Time1 second to within 2% of set point
- Controller Input Power Required±15 VDC (±2%) @ 0.28 Amps
- Heater Input Power Required±15 VDC @ 1.5 Amps (3 Amps with 90 - 150°C option)
- Set Point Signal0 - 5 VDC from <20 kOhm
- Flow Output Signal0 - 5 VDC into >10 kOhm
- Temperature Output10 mV/°C (0°C=0 VDC)
- Relay Contact Ratings2 Amps @ 28 VDC, 1 Amp @ 120 VAC resistive
- Signal Connector15-pin Type "D", RFI/EMI shielded
- Heater Connector9-pin Type "D", RFI/EMI shielded
- Maximum Line Pressure35 psia
- External Leak Integrity<1 x 10-9 scc/sec He
- Leak Integrity Through Closed Valve<1% of Full Scale or 1 sccm, whichever is greater
- Wetted MaterialsInconel, 316 Stainless Steel, Nickel
- Mounting PositionDo not mount upside-down
- Fitting TypeSwagelok® 8 VCR® male
- Weight7.9 lbs (3.59 kg)