Technical Specs

  • Type
    Vacuum/pressure Transducer
  • Measurement Type
    Absolute
  • Full Scale Pressure Ranges
    1-25,000 Torr (mmHg) (0.02-500 psia)
  • Output
    0-5 VDC or 0-10 VDC into >10 kΩ load
  • Accuracy
    0.5% of Reading
  • Zero Temperature Coefficient
    0.008% of Full Scale/°C (10 Torr through 25,000 Torr)
    0.020% of Full Scale (1 and 2 Torr)
  • Span Temperature Coefficient
    0.04% of Reading/°C
  • Operating Temperature
    0-50°C
    15-40°C (1 and 2 Torr only)
  • Overpressure Limit
    45 psia or 2x Full Scale, whichever is greater
  • Burst Pressure
    10x Full Scale or 100 psi, whichever is greater
  • Exposed Materials
    Inconel®
  • Compliance
    Fully CE Compliant to EMC Directive 2004/108/EC1
    Fully compliant to RoHS Directive 2002-95-EC

Technical Specs

  • Type
    Vacuum/pressure Transducer
  • Measurement Type
    Absolute
  • Full Scale Pressure Ranges
    1-25,000 Torr (mmHg) (0.02-500 psia)
  • Output
    0-5 VDC or 0-10 VDC into >10 kΩ load
  • Accuracy
    0.5% of Reading
  • Zero Temperature Coefficient
    0.008% of Full Scale/°C (10 Torr through 25,000 Torr)
    0.020% of Full Scale (1 and 2 Torr)
  • Span Temperature Coefficient
    0.04% of Reading/°C
  • Operating Temperature
    0-50°C
    15-40°C (1 and 2 Torr only)
  • Overpressure Limit
    45 psia or 2x Full Scale, whichever is greater
  • Burst Pressure
    10x Full Scale or 100 psi, whichever is greater
  • Exposed Materials
    Inconel®
  • Compliance
    Fully CE Compliant to EMC Directive 2004/108/EC1
    Fully compliant to RoHS Directive 2002-95-EC