Sorry, this product is no longer offered for sale. The recommended replacement product is 629h-heated-absolute-capacitance-manometers.
629F 0.02-25,000 Torr, Heated (45 or 100°C) Absolute Baratron® Capacitance Manometers
Model: 629f-heated-absolute-capacitance-manometers
629F 0.02-25,000 Torr, Heated (45 or 100°C) Absolute Baratron® Capacitance Manometers
Model: 629f-heated-absolute-capacitance-manometers

Overview
Overview
Technical Specs
- TypePressure Transducer
- Sensor TypeCapacitance Manometer
- Measurement TypeAbsolute
- Sensor TemperatureControlled to 45°C or 100°C
- Full Scale Pressure Ranges0.02 (45°C only), 0.05, 0.1, 0.25, 1, 2, 10, 20, 100, 1000, 2000, 5000, 10000, 15000, 20000, 25000 Torr
- Accuracy45°C version: 0.12% of reading for 1-25,000 Torr
0.15% of reading for 0.25, 0.1, and 0.05 Torr
0.25% for 0.02 Torr - Analog Output0 - 10 VDC into >10K Ω load
- Analog Connector9-pin D-sub
- Span Temperature Coefficient0.02% of Reading/°C
- Ambient Operating Temperature45°C version: 15° to 40°C
100°C version: 15° to 50°C - Volume6.3 cc
- Warm-up Time2 hours for 1 - 25,000 Torr ranges, 4 hours for ranges < 1 Torr
- Overpressure Limit45 psia (310 kPa) or 120% Full Scale, whichever is greater
- Materials Exposed to Process GasesInconel®
- Power Requirements±15 VDC ±5% @ 0.25 A (45°C) or 0.6 (100°C)
- ComplianceCE
Technical Specs
- TypePressure Transducer
- Sensor TypeCapacitance Manometer
- Measurement TypeAbsolute
- Sensor TemperatureControlled to 45°C or 100°C
- Full Scale Pressure Ranges0.02 (45°C only), 0.05, 0.1, 0.25, 1, 2, 10, 20, 100, 1000, 2000, 5000, 10000, 15000, 20000, 25000 Torr
- Accuracy45°C version: 0.12% of reading for 1-25,000 Torr
0.15% of reading for 0.25, 0.1, and 0.05 Torr
0.25% for 0.02 Torr - Analog Output0 - 10 VDC into >10K Ω load
- Analog Connector9-pin D-sub
- Span Temperature Coefficient0.02% of Reading/°C
- Ambient Operating Temperature45°C version: 15° to 40°C
100°C version: 15° to 50°C - Volume6.3 cc
- Warm-up Time2 hours for 1 - 25,000 Torr ranges, 4 hours for ranges < 1 Torr
- Overpressure Limit45 psia (310 kPa) or 120% Full Scale, whichever is greater
- Materials Exposed to Process GasesInconel®
- Power Requirements±15 VDC ±5% @ 0.25 A (45°C) or 0.6 (100°C)
- ComplianceCE
Resources & Downloads
Drawings & CADs
629 Baratron® Heated Absolute Capacitance Manometer Dimensional Drawing(85.8 kB, PDF) 629 Dimensional Drawing(126.9 kB, PDF)
Manuals
629D Baratron® Heated Absolute Capacitance Manometer Manual(504.2 kB, PDF) 629 and E29 Capacitance Manometer Pin-outs(21.8 kB, PDF)
Literature
629F Baratron® Heated Capacitance Manometer (archive)(181.4 kB, PDF)