Sorry, this product is no longer offered for sale. The recommended replacement product is 722C53TGA2FJ.
722B 1-25,000 Torr, General Purpose Absolute Baratron® Capacitance Manometer
Model: 722B53TGA2FJ
722B 1-25,000 Torr, General Purpose Absolute Baratron® Capacitance Manometer
Model: 722B53TGA2FJ

Overview
Overview
The 722B53TGA2FJ General Purpose Absolute Baratron® Capacitance Manometer features 5000 Torr Full Scale pressure measurement with ±0.5% of reading accuracy and a NW16-KF Flange. Electrical connections for the +13 to +32 VDC power requirements and 0-10 VDC proportional analog output are made with a 5-pin terminal strip connector. It features a flexible design making it an ideal solution for retrofit applications as well as new equipment designs. Robust sensor construction provides high overpressure tolerances, reducing errors caused by occasional line pressure spikes. All sensor wetted surfaces are made from Inconel® to ensure maximum compatibility with many process gases. Sophisticated technology and strict process controls ensure clean, smooth finishes (<10 Ra) on all wetted surfaces. Signal conditioning circuitry is assembled using the latest techniques in surface mount technology. All transducers are backed with a full two-year warranty.
Technical Specs
- TypePressure Transducer
- Measurement TypeAbsolute
- Full Scale Range5000 Torr
- Fitting TypeNW16 ISO-KF
- Analog Output0 - 10 VDC
- Electrical Connector9-pin D-sub
- Accuracy0.5% of Reading
- Response Time<20 msec
- Zero Temperature Coefficient0.008% of F.S./°C
- Span Temperature Coefficient0.04% of Rdg./°C
- Operating Temperature Range0 - 50°C
- Overpressure Limit45 psia or 2 x F.S., whichever is greater
- Burst Pressure10 x F.S. or 100 psi, whichever is greater
- Exposed MaterialsInconel®
- Power Requirements13 - 32 VDC
Technical Specs
- TypePressure Transducer
- Measurement TypeAbsolute
- Full Scale Range5000 Torr
- Fitting TypeNW16 ISO-KF
- Analog Output0 - 10 VDC
- Electrical Connector9-pin D-sub
- Accuracy0.5% of Reading
- Response Time<20 msec
- Zero Temperature Coefficient0.008% of F.S./°C
- Span Temperature Coefficient0.04% of Rdg./°C
- Operating Temperature Range0 - 50°C
- Overpressure Limit45 psia or 2 x F.S., whichever is greater
- Burst Pressure10 x F.S. or 100 psi, whichever is greater
- Exposed MaterialsInconel®
- Power Requirements13 - 32 VDC
Resources & Downloads
Manuals
722 Capacitance Manometer Pin-outs(20.5 kB, PDF)
Drawings & CADs