623F 1-1000 Torr, Unheated Absolute Baratron® Capacitance Manometers with Trip Points
Model: 623F11TDE
623F 1-1000 Torr, Unheated Absolute Baratron® Capacitance Manometers with Trip Points
Model: 623F11TDE

Overview
Overview
The 623F11TDE Unheated Absolute Baratron® Capacitance Manometer with Trip Points features 10 Torr full scale pressure measurement with 0.25% accuracy, and a size NW16 ISO-KF fitting. It has two user adjustable process trip point relays, each with an adjustment range from 0.1% to 100% of Full Scale or voltage equivalent 0.01 to 10 VDC. Dual color LEDs indicate relay status. Typical relay uses include pump cross-over and valve activation. The 623F is an ambient temperature transducer. The instrument operates with ±15 VDC (±5%) input at <100 mA, and provides an output of 0 to 10 VDC linear with pressure. The 623F's sensor exposes only Inconel® to the process permitting the use with corrosive or dirty gases. Measurements are independent of gas composition.
Technical Specs
- TypePressure Transducer
- Sensor TypeCapacitance Manometer
- Measurement TypeAbsolute
- Full Scale Range10 Torr
- Sensor TemperatureAmbient
- Fitting TypeNW16 ISO-KF
- Analog Output0 - 10 VDC
- Accuracy0.25% of Reading
- Electrical ConnectorTerminal block
- Zero Temperature Coefficient0.005% of Full Scale/°C
- Span Temperature Coefficient0.04% of Reading/°C
- Operating Temperature Range0 - 50°C
- Resolution0.001% of Full Scale
- Volume6.3 cc
- Overpressure Limit45 psia (310 kPa)
- Exposed MaterialsIncoloy® and Inconel®
- Warm-up Time30 minutes
- Trip Points2
- Trip Point Relay RatingSPDT, 1 Amp @ 30 VDC or 0.5 Amp @ 30 VAC resistive
- Power Requirements±15 VDC ±5% @ 75 mA
- ComplianceCE
Technical Specs
- TypePressure Transducer
- Sensor TypeCapacitance Manometer
- Measurement TypeAbsolute
- Full Scale Range10 Torr
- Sensor TemperatureAmbient
- Fitting TypeNW16 ISO-KF
- Analog Output0 - 10 VDC
- Accuracy0.25% of Reading
- Electrical ConnectorTerminal block
- Zero Temperature Coefficient0.005% of Full Scale/°C
- Span Temperature Coefficient0.04% of Reading/°C
- Operating Temperature Range0 - 50°C
- Resolution0.001% of Full Scale
- Volume6.3 cc
- Overpressure Limit45 psia (310 kPa)
- Exposed MaterialsIncoloy® and Inconel®
- Warm-up Time30 minutes
- Trip Points2
- Trip Point Relay RatingSPDT, 1 Amp @ 30 VDC or 0.5 Amp @ 30 VAC resistive
- Power Requirements±15 VDC ±5% @ 75 mA
- ComplianceCE