Specifications

  • Type
    Pressure Transducer
  • Sensor Type
    Industrial Capacitance Manometer
  • Measurement Type
    Differential
  • Full Scale Pressure Ranges
    20 to 3000 PSI (1.4 to 204 bar)
  • Accuracy
    ±1% of Reading
  • Resolution
    ±0.01% of Full Scale
  • Repeatability
    ±0.1% of Reading
  • Analog Output
    0 to 10 VDC into >10 k Ω load
  • Electrical Connector
    15-pin D-subminiture
  • Zero Temperature Coefficient
    0.02% of Full Scale/°C
  • Span Temperature Coefficient
    0.04% of Reading/°C
  • Volume
    4.7 cc
  • Power Requirements
    ±15VDC (± 5%) @ 35 mA (max)
  • Factory Default Trip Relay Settings
    Trip Relay A energizes below 55% of full-scale range, and Trip Relay B energizes above 55% of full-scale range.
  • Relay Specifications
    Two independently-adjusted DPDT contacts adjustable between 0.5% and 100% of FS. Relays are UL listed and rated at 1.0 amps @ 30 VDC or 0.3 amps @ 30 VAC.
  • Weight
    1.0 lbs (0.4 kg)
  • Compliance
    Compliant with EMC Directive 2004/108/EC
    Compliant with Directive 2002-95-EC

Features

15-pin Type "D" Pin-Outs

Pin Description
1 Relay Contact A, N.O.
2 Pressure Output
3 Reserved
4 Reserved
5 Power Return
6 -15 VDC
7 +15 VDC
8 Relay Contact A, N.C.
9 Relay Contact B, N.C.
10 Trip Point A, Setting
11 Trip Point B, Setting
12 Pressure Return
13 Relay contact B, N.O.
14 Relay Contact B, Common
15 Relay contact A, Common

Need help?